Fluke Calibration 1502A and 1504 Tweener

product id: 105791

Category: Thermometers

Manufacturer: Fluke Calibration

InstruHub is a leading supplier and reseller of Fluke Calibration Thermometers in UAE, and we ship to all major cities such as Dubai, Abu- Dhabi, Sharjah


Features

  • Two models to choose from reading PRTs or thermistors
  • Battery packs available
  • Best price/performance package

The Fluke Calibration Tweener (formerly Hart Calibration) thermometer readouts consists of two models - the 1502A and 1504. No other thermometer comes close to the performance and features of the Fluke Tweener thermometer readouts for anywhere near the price.

1502A Tweener PRT Readout

The 1502A readout features accuracy up to ±0.006°C. In addition, it reads 100-ohm, 25-ohm, and 10-ohm probes, has a resolution of 0.001°C across its entire range, and is the smallest unit in its class. The 1502A readout also has an optional battery pack for completely portable operation.

Each 1502A readout is programmable to match a probe's constants for maximum linearity and accuracy. All probe constants and coefficients are programmed through simple, front-panel keystrokes. Temperature is displayed in °C, °F, K, or resistance in ohms.

The 1502A readout accurately measures the resistance of the probe and then converts the resistance to a temperature value using its built-in algorithms.

For convenience, the 1502A readout reads the common industrial grade IEC-751 or "385" ALPHA RTD without any programming. Enter the actual R0 and ALPHA of the individual probe for increased accuracy. For maximum accuracy, use the ITS-90 formulas. The 1502A readout accepts the subranges 4 and 6 through 11.

ITS-90 formulas reside in the Tweener's firmware. If your probe has been calibrated for any of the above subranges of the ITS-90, you simply enter the coefficients directly into the 1502A readout.

Each thermometer comes complete with an RS-232 interface for automation of temperature data collection, calibrations, or process control functions. An IEEE-488 interface is available as an option.

The 1502A readout is calibrated digitally using the front-panel buttons. You never have to open the box to calibrate it. This calibration protocol further reduces the cost of the 1502A. It goes where you go and works the way you want it to.

1504 Tweener Thermistor Readout

If you need more accuracy in a limited temperature range, the 1504 readout gives it to you as a thermistor readout. Thermistors are less fragile than PRTs and less likely to be impacted by mechanical shock. Thermistors are more sensitive to temperature, have faster response times, and come in many shapes for different applications.

Typical accuracy of the 1504 is ±0.002°C with a resolution of 0.0001°C.

Why Buy from InstruHub:




Similar products

call
phone whatsApp mail